MATERIALS

Materials Provided

Test devices Droppers Buffer Package insert

Materials Required But Not Provided

Specimen collection containers Lancets (for fingerstick whole blood only)

Centrifuge Micropipette Timer

adv_img

DIRECTIONS FOR USE

1. Bring the pouch to room temperature before opening it. Remove the test device from the sealed pouch and use it as soon as possible. Best results will be obtained if the assay is performed within one hour.

2. Place the test device on a clean and level surface. Hold the dropper

vertically and transfer 1 free fall drop of serum or plasma(approximately 10 L)and 2 drops of buffer(approximately 80 L)to each specimen well(S) of the test device respectively, and then start the timer. Avoid trapping air bubbles in the specimen well. See illustration below.

3. Wait for the colored line(s) to appear. The results should be read at 10 minutes. Do not interpret the results after 15 minutes.

qwe